During product registration and changing registration, EMC test report and electrical safety test report shall be related to ensure that the electrical safety test sample is consistent with the EMC test sample.
During Registration Renewal, if only EMC performance is tested, it is not required to associate with the electrical safety testing report of the previous registration. Because after the product is approved and registered, the administrative counterparts shall organize production according to the approved certificate to keep the product unchanged from the previous registration, so it is not necessary to associate when renewing the certificate.
During the validity period of the registration certificate, if there is any change of the product that does not involve the Change of Permission Items, the administrative counterparts shall carry out relevant verification and validation according to the requirements of the quality management system to ensure that the change does not affect the safety and effectiveness of the product. In addition, when applying for Registration Renewal, it shall state in the statement of no change of products for Registration Renewal: “the change of products shall be controlled through the quality management system, and the items stated in the registration certificate shall not change.