What Aspects Should Be Considerded In Registration Submission When Multiple Receiving Coils In MRI System Are Used Jointly?


2022-10-30

In the case of joint use of multiple receiving coils in the MRI system, the following issues shall be considered in the registration submission:

  1. The combination mode, patient positioning mode and corresponding scanning position of the coil should be described in the overview material;
  2. The performance of SNR, uniformity, 2D scanning layer thickness, spatial resolution and ghost should be provided in the research data;
  3. The product technical requirements shall include the performance requirements of single coil and the performance requirements of multi coil joint use;

For clinical evaluation, the joint combined coil should be combined with the appropriate scanning position.